Producto | Equipo de mamografía Shenographe 600T Senix HF |
Empresa | Voxel Argentina, S.A. |
Categoria | Mamógrafos |
Marca | Otra |
Especialidad médica | Ginecología |
Tipo de operación | Venta |
Estado | Nuevo |
The Senographe 600T Senix HF system from GE Medical Systems provides the tecnology needed for high-quality images, clinical versatility and ease of use.
TUBE
Focal spot size: Dual focus 0.1/0.3 mm
Anode material: Molybdenum/Vanadium rotating anode.
Anode Angle: 0.1 focal spot: -91/4 0.3 focal spot: 01/4
Heat Loading: 90,000 HU air-cooled
Power: 0.1 - 1 kW0.3 - 55 kW
Maximun mA: 0.1 - 30 mA 0.3 - 100 mA
Filtration: Inherent: 0.8 mm Be
Added: Molybdenum = 0.03 mm for (equal or minor) 36 kV
Aluminium = 1.00 mm
for > 36 kV
Heat dissipation: 60,000 HU/minute
GENERATOR
Type: Single-phase, high-frequency, constant voltage
mAs range: 4 to 600 mAs, 27 steps
kV range: 22 to 49 kVp in kV increments
Exposure time: 0.04 to 10 sec.Automatic exposure control: Autocomp with microprocessor
control for five screen/film combinations
mAs meter: automatic exposure mode
Density steps: + - 5 density steps, each step 20%
CONTROL PANEL
The control panel is built into the cabinet and enables operation either:
Using Autocomp, the automatic exposure feature, and 1-point mode, i.e. with kV selection and mAs readout on the control panel.
Using 2-point mode operation, i.e. kV and mAs selection.
The following indicator lamps and audible tones are located in the control panel:
Main power supply.
Focal spot size selection.
Mo / Al filter selection.
kV selection (off when no diaphragm is in place).
mAs selection.
X-ray exposure lamp.
Autocomp automatic exposure control reset lamp
System on.
The pressure gauge and the pneumatic compression adjustment knob are positionedon the opposite side of the control panel. Emergency stop buttons are positioned on either side of the column; whenactivated, compression release is manual.
PNEUMATIC COMPRESSION
Compression is foot pedal-controlled to reduce operator constraint and ensure greater ease in patient positioning.
The pneumatic compression system, together with improved paddle fixation, is automatically adapted to match breast shape and density.
Compression is calibrated to automatically stop at a pre set compression force, eliminating unnecessary manual adjustments. Pressure is adjustable from 0 to 4 bars, producing a force between 0 and 20 Kg.